This report presents the results of the key comparison EURAMET.M.D-K4.2020 (EURAMET 1496) that covered the calibration of high-resolution hydrometers for liquid in the density range 600 kg/m3 to 2 000 kg/m3 at the reference temperature of 20 °C. Nine laboratories participated in this key comparison, eight European national metrology institutes and one from Asia Pacific metrology: NIM from China. The participating laboratories were initially divided into two petals linked by two density laboratories: INRIM and PTB. In each petal, four similar transfer standards, hydrometers in the density range 600 kg/m3 to 2 000 kg/m3 are used. The equipment used by the participants was found to have very similar characteristics. Most of the declared relative uncertainties were within the 5×10-5 . The results of all participated laboratories have been linked to the CIPM key comparison CCM.D-K4 by INRIM and PTB. Discrepant results have been reported for three laboratories, which, however, do not affect their CMCs.

EURAMET Key Comparison EURAMET.M.D-K4.2020: Hydrometer calibration comparison from 600 kg/m3 to 2000 kg/m3 / Malengo, Andrea; Eppers, Daniela; Akcadag, Umit Y; Furtado, Andreia; Schiebl, Markus; Vamossy, Csilla; Bebic, Jelena; Milutinovic, Milan; Jintao, Wang. - In: METROLOGIA. - ISSN 0026-1394. - 59:1A(2022), p. 07005. [10.1088/0026-1394/59/1A/07005]

EURAMET Key Comparison EURAMET.M.D-K4.2020: Hydrometer calibration comparison from 600 kg/m3 to 2000 kg/m3

Malengo, Andrea
;
2022

Abstract

This report presents the results of the key comparison EURAMET.M.D-K4.2020 (EURAMET 1496) that covered the calibration of high-resolution hydrometers for liquid in the density range 600 kg/m3 to 2 000 kg/m3 at the reference temperature of 20 °C. Nine laboratories participated in this key comparison, eight European national metrology institutes and one from Asia Pacific metrology: NIM from China. The participating laboratories were initially divided into two petals linked by two density laboratories: INRIM and PTB. In each petal, four similar transfer standards, hydrometers in the density range 600 kg/m3 to 2 000 kg/m3 are used. The equipment used by the participants was found to have very similar characteristics. Most of the declared relative uncertainties were within the 5×10-5 . The results of all participated laboratories have been linked to the CIPM key comparison CCM.D-K4 by INRIM and PTB. Discrepant results have been reported for three laboratories, which, however, do not affect their CMCs.
File in questo prodotto:
File Dimensione Formato  
EURAMET 1496 Draft B October 2021 accepted r1-1.pdf

accesso aperto

Tipologia: final published article (publisher’s version)
Licenza: Pubblico - Tutti i diritti riservati
Dimensione 593.69 kB
Formato Adobe PDF
593.69 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11696/76167
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact