The technical report summarizes the results provided by the Resistance laboratory of the National Research Institute (INRIM) concerning the calibration of two 1 Ω and two 10 kΩ standard travelling resistors related to the comparison BIPM.EM-K13.a (1 Ω resistance standards) and BIPM.EM-K13.b (10000 Ω resistance standards) which constitute the BIPM programme of on-going bilateral comparisons of resistance standards. The chosen resistors normally assure a transfer uncertainty less than a few parts in 108. The comparison was carried out following an initial/return measurement pattern: the BIPM travelling standards were first calibrated at the BIPM then they were sent to INRIM where they were calibrated, and finally re-calibrated after their return at the BIPM. All measurement results will be corrected to the reference temperature (23.000 °C) and the reference pressure (1013.25 hPa). The result of the comparison will be expressed as the difference between the calibrations from the two laboratories, together with its associated uncertainty. The pilot laboratory was the BIPM.

NRIM-BIPM.EM-K13.a and b Key Comparisons of Electrical Resistance Standards using 1 Ω and 10 kΩ Resistors as Travelling Standards: INRIM results / Capra, PIER PAOLO; Galliana, Flavio; Francese, Claudio; Lanzillotti, Marco; RONCAGLIONE TET, Luca. - (2023), pp. 1-12.

NRIM-BIPM.EM-K13.a and b Key Comparisons of Electrical Resistance Standards using 1 Ω and 10 kΩ Resistors as Travelling Standards: INRIM results

Pier Paolo Capra
Methodology
;
Flavio Galliana
Data Curation
;
Claudio Francese
Membro del Collaboration Group
;
Marco Lanzillotti
Membro del Collaboration Group
;
Luca Roncaglione Tet
Membro del Collaboration Group
2023

Abstract

The technical report summarizes the results provided by the Resistance laboratory of the National Research Institute (INRIM) concerning the calibration of two 1 Ω and two 10 kΩ standard travelling resistors related to the comparison BIPM.EM-K13.a (1 Ω resistance standards) and BIPM.EM-K13.b (10000 Ω resistance standards) which constitute the BIPM programme of on-going bilateral comparisons of resistance standards. The chosen resistors normally assure a transfer uncertainty less than a few parts in 108. The comparison was carried out following an initial/return measurement pattern: the BIPM travelling standards were first calibrated at the BIPM then they were sent to INRIM where they were calibrated, and finally re-calibrated after their return at the BIPM. All measurement results will be corrected to the reference temperature (23.000 °C) and the reference pressure (1013.25 hPa). The result of the comparison will be expressed as the difference between the calibrations from the two laboratories, together with its associated uncertainty. The pilot laboratory was the BIPM.
2023
3/2023
open
File in questo prodotto:
File Dimensione Formato  
RT-2023-03.pdf

accesso aperto

Descrizione: pdf
Tipologia: Versione editoriale
Licenza: Pubblico - Tutti i diritti riservati
Dimensione 1.03 MB
Formato Adobe PDF
1.03 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11696/75319
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact