This paper addresses the calibration of lock-in amplifiers in the low frequency range. A simple but effective calibration setup implemented at INRIM is described. This includes a stable low-voltage source, composed of a digital sine wave generator and a cascade of two voltage dividers, one inductive and one resistive. The magnitude error is the lock-in magnitude error, which can be determined with an uncertainty of the order of 10(-4) or better, that is, two orders of magnitude lower than the typical lock-in accuracy specification. The system can operate with an input voltage from 0.1 mu V to 10 mu V and a frequency from a few hertz to about 1 kHz. The performances of the calibration setup were evaluated at 1 mu V and 10 mu V at 20 Hz and 103 Hz and compared with those of another existing setup, implemented at METAS, through a direct comparison in the framework of an EURAMET research project.

A new calibration setup for lock-in amplifiers in the low frequency range and its validation in a bilateral comparison / Cultrera, Alessandro; Corminboeuf, David; D’Elia, Vincenzo; Tran, Ngoc Thanh Mai; Callegaro, Luca; Ortolano, Massimo. - In: METROLOGIA. - ISSN 0026-1394. - 58:2(2021), p. 025001. [10.1088/1681-7575/abdae0]

A new calibration setup for lock-in amplifiers in the low frequency range and its validation in a bilateral comparison

Cultrera, Alessandro
;
D’Elia, Vincenzo;Callegaro, Luca;Ortolano, Massimo
2021

Abstract

This paper addresses the calibration of lock-in amplifiers in the low frequency range. A simple but effective calibration setup implemented at INRIM is described. This includes a stable low-voltage source, composed of a digital sine wave generator and a cascade of two voltage dividers, one inductive and one resistive. The magnitude error is the lock-in magnitude error, which can be determined with an uncertainty of the order of 10(-4) or better, that is, two orders of magnitude lower than the typical lock-in accuracy specification. The system can operate with an input voltage from 0.1 mu V to 10 mu V and a frequency from a few hertz to about 1 kHz. The performances of the calibration setup were evaluated at 1 mu V and 10 mu V at 20 Hz and 103 Hz and compared with those of another existing setup, implemented at METAS, through a direct comparison in the framework of an EURAMET research project.
2021
partially_open
File in questo prodotto:
File Dimensione Formato  
20201117_LockinComparison_Metrologia.pdf

accesso aperto

Tipologia: Documento in Pre-print
Licenza: Pubblico - Tutti i diritti riservati
Dimensione 7 MB
Formato Adobe PDF
7 MB Adobe PDF Visualizza/Apri
Cultrera_2021_Metrologia_58_025001.pdf

non disponibili

Tipologia: Versione editoriale
Licenza: Non Pubblico - Accesso privato/ristretto
Dimensione 1.54 MB
Formato Adobe PDF
1.54 MB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11696/72974
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 7
  • ???jsp.display-item.citation.isi??? 5
social impact