High sensitive triaxial accelerometers are used in several space missions to measure the non-gravitational accelerations acting on the spacecraft. Among these, the capacitive accelerometers developed for ESA missions Jupiter Icy moons Explorer (JUICE) and BepiColombo were designed to measure accelerations of the order of 3 · 10−6 m s−2 with an accuracy level of 300 ppm in the frequency range (3 · 10−5–0.1) Hz. Despite the signal to be measured is of the same order of magnitude of the seismic noise on the earth, an accurate on-ground calibration is needed. The facility set-up at INRIM to this purpose is based on a simple principle: the base of the accelerometer is dynamically tilted by an angle α so that the sensor undergoes a component of the gravitational acceleration g proportional to angle α. In practice, several issues have to be addressed by the calibration facility, such as the seismic noise limiting the signal-to-noise ratio, the generation and the accurate measurement of the tilt angle. Furthermore, the calibration was performed taking into account different on-flight conditions such as different operating temperatures and possible deformation of the accelerometers during the launch. The experimental set-up and the calibration procedure are described in the paper. The measurement results and the uncertainty budget show that a relative accuracy of 240 ppm has been achieved.
Calibration of high accuracy accelerometers for ESA missions BepiColombo and JUICE at INRIM / Astrua, Milena; Pisani, Marco; Santiano, Marco; Audrito, Emanuele. - In: MEASUREMENT SCIENCE & TECHNOLOGY. - ISSN 1361-6501. - 34:10(2023). [10.1088/1361-6501/ace20c]
Calibration of high accuracy accelerometers for ESA missions BepiColombo and JUICE at INRIM
Milena Astrua
;Marco Pisani;Marco Santiano;Emanuele Audrito
2023
Abstract
High sensitive triaxial accelerometers are used in several space missions to measure the non-gravitational accelerations acting on the spacecraft. Among these, the capacitive accelerometers developed for ESA missions Jupiter Icy moons Explorer (JUICE) and BepiColombo were designed to measure accelerations of the order of 3 · 10−6 m s−2 with an accuracy level of 300 ppm in the frequency range (3 · 10−5–0.1) Hz. Despite the signal to be measured is of the same order of magnitude of the seismic noise on the earth, an accurate on-ground calibration is needed. The facility set-up at INRIM to this purpose is based on a simple principle: the base of the accelerometer is dynamically tilted by an angle α so that the sensor undergoes a component of the gravitational acceleration g proportional to angle α. In practice, several issues have to be addressed by the calibration facility, such as the seismic noise limiting the signal-to-noise ratio, the generation and the accurate measurement of the tilt angle. Furthermore, the calibration was performed taking into account different on-flight conditions such as different operating temperatures and possible deformation of the accelerometers during the launch. The experimental set-up and the calibration procedure are described in the paper. The measurement results and the uncertainty budget show that a relative accuracy of 240 ppm has been achieved.File | Dimensione | Formato | |
---|---|---|---|
Astrua_2023_Meas._Sci._Technol._34_104001.pdf
accesso aperto
Descrizione: articolo
Tipologia:
final published article (publisher’s version)
Licenza:
Creative Commons
Dimensione
1.67 MB
Formato
Adobe PDF
|
1.67 MB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.