Current document reports on the development and validation of advanced practical models for describing force measuring devices used for the measurement of static, continuous forces in part A as well as dynamic forces in part B of the report. In the force measurement applications, the sensitivity curve of the force measuring devices determined in the calibration gives the traceability to national standards. The measurement uncertainty in an application is mainly affected by the sensitivity stability, the influence of temperature and other parasitic components. To capture these effects an advanced model for static and continuous forces was developed in part A which describes these influences especially in respect of their contribution to the measurement uncertainty. In contrast to the static calibration procedure, the dynamic calibration of the sensor has been always challenging because of several reasons such as the sophisticated nature of dynamic measurement, insufficient structural equipment, and parasitic effects which yield higher measurement uncertainty in comparison to static measurements and therefore inappropriate characterization of the force transducers. To close this knowledge gap in the dynamic calibration, the advanced model for dynamic forces was developed in part B. To consider the dynamic properties of force measuring devices the frequency dependency of the sensitivity and the influence of parasitic influences from temperature of the force measuring devices was considered and described. The target uncertainty is 1 % for up to 100 Hz and the target uncertainty is 2 % for higher frequency range from 100 Hz up to 1000 Hz.

Validation report which details the advanced models developed to describe a) static and continuous and b) dynamic force transfer standards taking into account sensitivity stability, temperature and other parasitic influences on the measurement uncertainty (target uncertainty is 1 % up to 100 Hz and 2 % between 100 - 1000 Hz) / Sander, J.; Mirian, D.; Frank, H.; Hassan, S.; Kumme, R.; Prato, A.; Germak, A.; Wozniak, M.; Izquierdo, G.; Carmen, M.; Vavrecka, L.; Dizdar, H.; Aydemir, B.; Korhonen, J.; Oliveira, R. S.; Hiti, M.. - (2022), pp. 1-80. [10.5281/zenodo.7113187]

Validation report which details the advanced models developed to describe a) static and continuous and b) dynamic force transfer standards taking into account sensitivity stability, temperature and other parasitic influences on the measurement uncertainty (target uncertainty is 1 % up to 100 Hz and 2 % between 100 - 1000 Hz)

Prato A.;Germak A.;
2022

Abstract

Current document reports on the development and validation of advanced practical models for describing force measuring devices used for the measurement of static, continuous forces in part A as well as dynamic forces in part B of the report. In the force measurement applications, the sensitivity curve of the force measuring devices determined in the calibration gives the traceability to national standards. The measurement uncertainty in an application is mainly affected by the sensitivity stability, the influence of temperature and other parasitic components. To capture these effects an advanced model for static and continuous forces was developed in part A which describes these influences especially in respect of their contribution to the measurement uncertainty. In contrast to the static calibration procedure, the dynamic calibration of the sensor has been always challenging because of several reasons such as the sophisticated nature of dynamic measurement, insufficient structural equipment, and parasitic effects which yield higher measurement uncertainty in comparison to static measurements and therefore inappropriate characterization of the force transducers. To close this knowledge gap in the dynamic calibration, the advanced model for dynamic forces was developed in part B. To consider the dynamic properties of force measuring devices the frequency dependency of the sensitivity and the influence of parasitic influences from temperature of the force measuring devices was considered and described. The target uncertainty is 1 % for up to 100 Hz and the target uncertainty is 2 % for higher frequency range from 100 Hz up to 1000 Hz.
File in questo prodotto:
File Dimensione Formato  
ComTraForce_deliverable_D2.pdf

accesso aperto

Descrizione: Deliverable2
Tipologia: Versione editoriale
Licenza: Creative Commons
Dimensione 4.84 MB
Formato Adobe PDF
4.84 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11696/75840
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact