We used absolute primary acoustic gas thermometry (AGT) to calibrate a Pt-Co resistance thermometer on the thermodynamic temperature scale by measuring the speed of sound in helium at a temperature T* chosen to be near the temperature of the triple point of neon, T (Ne). Prior to the present AGT, the Pt-Co thermometer was used with a neon triple-point cell as part of an interlaboratory comparison. Taken together, the results of the interlaboratory comparison and the present AGT redetermined the thermodynamic temperature T (Ne) = (24.555 15 +/- 0.000 24) K. This new value of T (Ne) is consistent with other recent determinations obtained with various primary methods. After completing the AGT thermodynamic calibration, we used the Pt-Co thermometer to link T* to the temperature ratios measured by single-pressure refractive-index gas thermometry (SPRIGT) in a different laboratory. (Gao et al 2020 Metrologia 57 065006) Now, the T*-linked SPRIGT system can calibrate other thermometers on the thermodynamic temperature scale T in the range 5 K <= T <= T (Ne) without using the international temperature scale ITS-90. At most temperatures in this range, the uncertainties of the T*-linked SPRIGT system are smaller than those of the ITS-90 systems used by National Metrology Institutes to calibrate resistance thermometers.

Acoustic measurement of the triple point of neon TNe and thermodynamic calibration of a transfer standard for accurate cryogenic thermometry / Pan, Changzhao; Sparasci, Fernando; Zhang, Haiyang; Gambette, Pascal; Plimmer, Mark; Imbraguglio, Dario; Gavioso, Roberto M; Moldover, Michael R; Gao, Bo; Pitre, Laurent. - In: METROLOGIA. - ISSN 0026-1394. - 58:4(2021), p. 045006. [10.1088/1681-7575/ac0711]

Acoustic measurement of the triple point of neon TNe and thermodynamic calibration of a transfer standard for accurate cryogenic thermometry

Imbraguglio, Dario;Gavioso, Roberto M;
2021

Abstract

We used absolute primary acoustic gas thermometry (AGT) to calibrate a Pt-Co resistance thermometer on the thermodynamic temperature scale by measuring the speed of sound in helium at a temperature T* chosen to be near the temperature of the triple point of neon, T (Ne). Prior to the present AGT, the Pt-Co thermometer was used with a neon triple-point cell as part of an interlaboratory comparison. Taken together, the results of the interlaboratory comparison and the present AGT redetermined the thermodynamic temperature T (Ne) = (24.555 15 +/- 0.000 24) K. This new value of T (Ne) is consistent with other recent determinations obtained with various primary methods. After completing the AGT thermodynamic calibration, we used the Pt-Co thermometer to link T* to the temperature ratios measured by single-pressure refractive-index gas thermometry (SPRIGT) in a different laboratory. (Gao et al 2020 Metrologia 57 065006) Now, the T*-linked SPRIGT system can calibrate other thermometers on the thermodynamic temperature scale T in the range 5 K <= T <= T (Ne) without using the international temperature scale ITS-90. At most temperatures in this range, the uncertainties of the T*-linked SPRIGT system are smaller than those of the ITS-90 systems used by National Metrology Institutes to calibrate resistance thermometers.
File in questo prodotto:
File Dimensione Formato  
Pan_at_al_AGT_25K_Metrologia_2021.pdf

accesso aperto

Tipologia: Versione editoriale
Licenza: Creative Commons
Dimensione 2.84 MB
Formato Adobe PDF
2.84 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11696/73572
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 5
  • ???jsp.display-item.citation.isi??? 6
social impact