ISO 4545-2 and 4545-3 of Knoop hardness tests require the geometrical verification of the indenters.INRiM hardness laboratory, in cooperation with Galileo-LTF, has developed the Gal-Indent optical mea-suring system for the verification of Vickers indenters. This system can measure the vertex anglesbetween two opposite faces, the quadrilateral base angles and the pyramid axis tilt angle. Using thesemeasured quantities as inputs of a suitable geometrical model, the angles between the opposite edgesat the vertex of Knoop indenters, nominally 172.5°and 130°, and the angle between the pyramid andindenter holder axes can be verified with an expanded uncertainty of 0.05°. Comparison of experimentalmeasurements performed on three different Knoop indenters, previously verified by an accredited labo-ratory, shows compatible results. The proposed geometrical model could be easily implemented by lab-oratories that adopt similar measuring systems addressed for the verification of Vickers indenterswithout any modification of the experimental apparatus
Verification of Knoop indenters with a Vickers-addressed optical system / Prato, Andrea; Origlia, Claudio; Germak, Alessandro. - In: MEASUREMENT. - ISSN 0263-2241. - 163:(2020), p. art. no. 107928. [10.1016/j.measurement.2020.107928]
Verification of Knoop indenters with a Vickers-addressed optical system
Prato, Andrea
Writing – Original Draft Preparation
;Origlia, ClaudioMembro del Collaboration Group
;Germak, AlessandroSupervision
2020
Abstract
ISO 4545-2 and 4545-3 of Knoop hardness tests require the geometrical verification of the indenters.INRiM hardness laboratory, in cooperation with Galileo-LTF, has developed the Gal-Indent optical mea-suring system for the verification of Vickers indenters. This system can measure the vertex anglesbetween two opposite faces, the quadrilateral base angles and the pyramid axis tilt angle. Using thesemeasured quantities as inputs of a suitable geometrical model, the angles between the opposite edgesat the vertex of Knoop indenters, nominally 172.5°and 130°, and the angle between the pyramid andindenter holder axes can be verified with an expanded uncertainty of 0.05°. Comparison of experimentalmeasurements performed on three different Knoop indenters, previously verified by an accredited labo-ratory, shows compatible results. The proposed geometrical model could be easily implemented by lab-oratories that adopt similar measuring systems addressed for the verification of Vickers indenterswithout any modification of the experimental apparatusFile | Dimensione | Formato | |
---|---|---|---|
1-s2.0-S0263224120304668-main.pdf
non disponibili
Tipologia:
final published article (publisher’s version)
Licenza:
Non Pubblico - Accesso privato/ristretto
Dimensione
1.3 MB
Formato
Adobe PDF
|
1.3 MB | Adobe PDF | Visualizza/Apri Richiedi una copia |
Verification of Knoop Indenters_post print.pdf
Open Access dal 19/05/2022
Tipologia:
accepted manuscript (author’s post-print)
Licenza:
Creative Commons
Dimensione
717.01 kB
Formato
Adobe PDF
|
717.01 kB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.