At the National Institute of Metrological Research (INRIM) a measuring setup able to vary, accurately and in a selective way, the temperature, humidity and pressure to detect eventual hysteretic phenomena and to evaluate their effects on electrical standards as high precision standard resistors and Zener-diode based DC Voltage standards. Main element of the setup is a pressure-tight case housing the standards and acting the stresses due to the desired parameters. By means of pressure-tight-connectors, the standards under test are connected to the measuring instruments, while the climatic conditions are acquired through inner and external sensors. Humidity is controlled by means of two conditioning containers filled respectively with water and silica-gel, which regulate the air-moisture percentage the case in a range of 10-90%. A pumping system establishes the pressure in the case from 600 hPa to 1400 hPa. The characterization vs. humidity and pressure is made at constant temperature by means of a commercial air-bath acting from 15 and 45 ° C. Preliminary results have been obtained for pressure dependence of commercial 100 Ω resistors and of an INRIM-developed 10 V Zener-diode based DC Voltage standard.

Measurement setup to characterize electrical standards / Capra, Pier Paolo; Cerri, Roberto; Galliana, Flavio; Lanzillotti, Marco; Pollarolo, Alessio; Roccato, Paolo. - (2019), p. 11003. (Intervento presentato al convegno 19th International Congress of Metrology tenutosi a PARIGI nel settembre 2019) [10.1051/metrology/201911003].

Measurement setup to characterize electrical standards

Capra, Pier Paolo;Cerri, Roberto;Galliana, Flavio
;
Lanzillotti, Marco;Roccato, Paolo
2019

Abstract

At the National Institute of Metrological Research (INRIM) a measuring setup able to vary, accurately and in a selective way, the temperature, humidity and pressure to detect eventual hysteretic phenomena and to evaluate their effects on electrical standards as high precision standard resistors and Zener-diode based DC Voltage standards. Main element of the setup is a pressure-tight case housing the standards and acting the stresses due to the desired parameters. By means of pressure-tight-connectors, the standards under test are connected to the measuring instruments, while the climatic conditions are acquired through inner and external sensors. Humidity is controlled by means of two conditioning containers filled respectively with water and silica-gel, which regulate the air-moisture percentage the case in a range of 10-90%. A pumping system establishes the pressure in the case from 600 hPa to 1400 hPa. The characterization vs. humidity and pressure is made at constant temperature by means of a commercial air-bath acting from 15 and 45 ° C. Preliminary results have been obtained for pressure dependence of commercial 100 Ω resistors and of an INRIM-developed 10 V Zener-diode based DC Voltage standard.
2019
19th International Congress of Metrology
settembre 2019
PARIGI
open
File in questo prodotto:
File Dimensione Formato  
metrology_cim2019_11003.pdf

accesso aperto

Descrizione: pdf
Tipologia: Versione editoriale
Licenza: Pubblico - Tutti i diritti riservati
Dimensione 776.44 kB
Formato Adobe PDF
776.44 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11696/60707
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact