In the Force Key Comparison CCM.F-K3 the measurand force was compared at the two force steps 500 kN and 1 MN. 12 laboratories participated in this comparison which was organised by PTB as the pilot laboratory in two laboratory groups (group A and B). In group A, the comparison was carried out with two 1 MN compression force transducers at the two force steps 500 kN and 1 MN (CCM.F-K3a) and with 6 participating laboratories. In group B, the comparison was carried out with two 500 kN compression force transducers at one force step of 500 kN (CCM.F-K3b) and with 9 participating laboratories. The key comparison reference values were determined as the weighted mean of all results for the two force steps and set to 500 kN and 1 MN, respectively, with the associated uncertainties. The degrees of equivalence were determined for all 12 laboratories for 500 kN compression force and for 6 laboratories for 1 MN compression force.
Final report on the force key comparison CCM.F-K3 / Kumme, Rolf; Averlant, Philippe; Bartel, Tom; Germak, ALESSANDRO FRANCO LIDIA; Knott, Andy; Man, John; Medina, Nieves; Ostrivnoy, Alexander; Park, You-Kyu; Röske, Dirk; Seifarth, Rick L.; Wozniak, Mikolaj; Ueda, Kazunaga; Zhimin, Zhang. - In: METROLOGIA. - ISSN 1681-7575. - 56:Number 1A, Technical Supplement 2019(2019), pp. 1-27. [10.1088/0026-1394/56/1A/07001]
Final report on the force key comparison CCM.F-K3
Alessandro Germak
Membro del Collaboration Group
;
2019
Abstract
In the Force Key Comparison CCM.F-K3 the measurand force was compared at the two force steps 500 kN and 1 MN. 12 laboratories participated in this comparison which was organised by PTB as the pilot laboratory in two laboratory groups (group A and B). In group A, the comparison was carried out with two 1 MN compression force transducers at the two force steps 500 kN and 1 MN (CCM.F-K3a) and with 6 participating laboratories. In group B, the comparison was carried out with two 500 kN compression force transducers at one force step of 500 kN (CCM.F-K3b) and with 9 participating laboratories. The key comparison reference values were determined as the weighted mean of all results for the two force steps and set to 500 kN and 1 MN, respectively, with the associated uncertainties. The degrees of equivalence were determined for all 12 laboratories for 500 kN compression force and for 6 laboratories for 1 MN compression force.File | Dimensione | Formato | |
---|---|---|---|
CCM.F-K3.pdf
accesso aperto
Tipologia:
final published article (publisher’s version)
Licenza:
Pubblico - Tutti i diritti riservati
Dimensione
792.63 kB
Formato
Adobe PDF
|
792.63 kB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.