The calibration of a multicomponent force transducers (MFTs)represents a challenge in the meganewtonrange. In fact,the generation of transversal forces and moments iscomplex since a force standard machine (FSM) is only able to apply an uniaxial force. Furthermore since MFTs are composed of multi-transducers, each one dedicated to a particular component, correlations between force and moment componentsare possible. Therefore, acalibration systemthat could simultaneously generate all force/moment components and could be suitableinevery FSMis needed. For this purpose, a coupleof tilted plates was designed. Calibration measurements were performed on a 2 MN MFT at INRiM, LNE and PTB. Exploitation matrixes and performance indicators showed good results, unless small but not negligible correlations between MFT outputs. In particularsome spurious valuesdue to the uncertainty inthe vertical force application pointinfluencedthe moment components.
Multicomponent force transducer calibration procedure using tilted plates / Palumbo, Stefano; Prato, Andrea; Mazzoleni, Fabrizio; Germak, ALESSANDRO FRANCO LIDIA. - In: JOURNAL OF PHYSICS. CONFERENCE SERIES. - ISSN 1742-6596. - 1065:042015(2018). [10.1088/1742-6596/1065/4/042015]
Multicomponent force transducer calibration procedure using tilted plates
Andrea PratoMembro del Collaboration Group
;Fabrizio MazzoleniData Curation
;Alessandro Germak
Methodology
2018
Abstract
The calibration of a multicomponent force transducers (MFTs)represents a challenge in the meganewtonrange. In fact,the generation of transversal forces and moments iscomplex since a force standard machine (FSM) is only able to apply an uniaxial force. Furthermore since MFTs are composed of multi-transducers, each one dedicated to a particular component, correlations between force and moment componentsare possible. Therefore, acalibration systemthat could simultaneously generate all force/moment components and could be suitableinevery FSMis needed. For this purpose, a coupleof tilted plates was designed. Calibration measurements were performed on a 2 MN MFT at INRiM, LNE and PTB. Exploitation matrixes and performance indicators showed good results, unless small but not negligible correlations between MFT outputs. In particularsome spurious valuesdue to the uncertainty inthe vertical force application pointinfluencedthe moment components.File | Dimensione | Formato | |
---|---|---|---|
Multicomponent force transducer calibration procedure using tilted plates.pdf
accesso aperto
Descrizione: paper
Tipologia:
final published article (publisher’s version)
Licenza:
Creative Commons
Dimensione
896.87 kB
Formato
Adobe PDF
|
896.87 kB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.