A portable temperature controlled DC Voltage and DC Resistance Reference with Switching Unit (RSU) has been developed at National Institute of Metrological Research (INRIM) to calibrate multifunction electrical instruments involving 10 V, a 1 Ω and a 10 kΩ Standards. The resistors are made with two resistors nets while the 10 V is a low noise-drift INRIM-projected circuit. Preliminary measurement over one-month period shows standards uncertainties span from 3.0×10–7 to 6.4×10–7, suitable for artifact calibration. With a real clock calendar, the RSU can show both the RSU Standards calibration values and updated ones between two calibrations according to an internal algorithm.
Portable DC Voltage and Resistance Reference with Switching Unit for Calibration of Electrical Precision Multifunction Instruments / Capra, PIER PAOLO; Galliana, Flavio; Cerri, Roberto; Lanzillotti, Marco; Pollarolo, Alessio. - (2018). ((Intervento presentato al convegno CONFERENCE ON PRECISE ELECTROMAGNETIC MEASUREMENTS tenutosi a PARIS nel JULY 2018.
Titolo: | Portable DC Voltage and Resistance Reference with Switching Unit for Calibration of Electrical Precision Multifunction Instruments |
Autori: | |
Data di pubblicazione: | 2018 |
Citazione: | Portable DC Voltage and Resistance Reference with Switching Unit for Calibration of Electrical Precision Multifunction Instruments / Capra, PIER PAOLO; Galliana, Flavio; Cerri, Roberto; Lanzillotti, Marco; Pollarolo, Alessio. - (2018). ((Intervento presentato al convegno CONFERENCE ON PRECISE ELECTROMAGNETIC MEASUREMENTS tenutosi a PARIS nel JULY 2018. |
Abstract: | A portable temperature controlled DC Voltage and DC Resistance Reference with Switching Unit (RSU) has been developed at National Institute of Metrological Research (INRIM) to calibrate multifunction electrical instruments involving 10 V, a 1 Ω and a 10 kΩ Standards. The resistors are made with two resistors nets while the 10 V is a low noise-drift INRIM-projected circuit. Preliminary measurement over one-month period shows standards uncertainties span from 3.0×10–7 to 6.4×10–7, suitable for artifact calibration. With a real clock calendar, the RSU can show both the RSU Standards calibration values and updated ones between two calibrations according to an internal algorithm. |
Handle: | http://hdl.handle.net/11696/58545 |
Appare nelle tipologie: | 4.1 Contributo in Atti di convegno |
File in questo prodotto:
File | Descrizione | Tipologia | Licenza | |
---|---|---|---|---|
CPEM-S401-10-1801261129.pdf | Versione editoriale | Non Pubblico - Accesso privato/ristretto | Utenti riconosciuti Richiedi una copia |