By agreement at the 10th WGFF meeting, the international comparison CCM.FF-K4.1.2011, for volume of liquids at 20 L and 100 mL, was performed during 2012–2014. Specially designed stainless steel pipettes were used as transfer standards for 20 L, whereas commercially available pycnometers were used for 100 mL. No discrepant measurements were distinguished on the 20 L artifacts. The largest difference between two NMIs was 0.0042 %, whereas the average degrees of equivalence for artifacts 710-04 and 710-05 resulted in 0.0001 % and 0.0005 %, respectively. Only one participant produced anomalous results for 100 mL measurements and results for artifacts 03.01.16 and 03.01.17 were all fully consistent with each other. The average degrees of equivalence for artifacts 03.01.16 and 03.01.17 were 0.00017 % and 0.0011 %, respectively.

By agreement at the 10th WGFF meeting, the international comparison CCM. FF-K4.1.2011, for volume of liquids at 20 L and 100 mL, was performed during 2012-2014. Specially designed stainless steel pipettes were used as transfer standards for 20 L, whereas commercially available pycnometers were used for 100 mL. No discrepant measurements were distinguished on the 20 L artifacts. The largest difference between two NMIs was 0.0042 %, whereas the average degrees of equivalence for artifacts 710-04 and 710-05 resulted in 0.0001 % and 0.0005 %, respectively. Only one participant produced anomalous results for 100 mL measurements and results for artifacts 03.01.16 and 03.01.17 were all fully consistent with each other. The average degrees of equivalence for artifacts 03.01.16 and 03.01.17 were 0.00017 % and 0.0011 %, respectively.

BIPM/CIPM key comparison CCM.FF-K4.1.2011. Final report for volume of liquids at 20 L and 100 mL / Arias, R; Maldonado, M; Batista, E; Jintao, W; Malengo, A; Malta, D; Ondodro, D; Penttinen, O; Smits, E; Wright, J. - In: METROLOGIA. - ISSN 1681-7575. - 52:07011(2015), pp. 07011-07011. [10.1088/0026-1394/52/1A/07011]

BIPM/CIPM key comparison CCM.FF-K4.1.2011. Final report for volume of liquids at 20 L and 100 mL

Malengo, A;
2015

Abstract

By agreement at the 10th WGFF meeting, the international comparison CCM. FF-K4.1.2011, for volume of liquids at 20 L and 100 mL, was performed during 2012-2014. Specially designed stainless steel pipettes were used as transfer standards for 20 L, whereas commercially available pycnometers were used for 100 mL. No discrepant measurements were distinguished on the 20 L artifacts. The largest difference between two NMIs was 0.0042 %, whereas the average degrees of equivalence for artifacts 710-04 and 710-05 resulted in 0.0001 % and 0.0005 %, respectively. Only one participant produced anomalous results for 100 mL measurements and results for artifacts 03.01.16 and 03.01.17 were all fully consistent with each other. The average degrees of equivalence for artifacts 03.01.16 and 03.01.17 were 0.00017 % and 0.0011 %, respectively.
File in questo prodotto:
File Dimensione Formato  
CCM.FF-K4.1.2011.pdf

solo utenti autorizzati

Tipologia: final published article (publisher’s version)
Licenza: Non Pubblico - Accesso privato/ristretto
Dimensione 862.53 kB
Formato Adobe PDF
862.53 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11696/52717
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? 2
social impact