During November 2011 a EURAMET key comparison of absolute gravimeters was organized in the Underground Laboratory for Geodynamics in Walferdange, Luxemburg. The comparison assembled 22 participants coming from 16 countries and four different continents. The comparison was divided into two parts: a key comparison that included six National Metrology Institutes or Designated Institutes, and a pilot study including all participants. The global result given by the pilot study confirms that all instruments are absolutely coherent with each other. The results obtained in the key comparison confirm a good agreement between the NMI instruments. Finally, a link to ICAG-2009 shows also that the NMI gravimeters are stable in time.

Final report of the regional key comparison EURAMET.M.G-K1: European Comparison of Absolute Gravimeters ECAG-2011 / O., Francis; G., Klein; H., Baumann; N., Dando; R., Tracey; C., Ullrich; S., Castelein; H., Hua; W., Kang; S., Chongyang; X., Songbo; T., Hongbo; L., Zhengyuan; V., Pálinkás; J., Kostelecký; J., Mäkinen; J., Näränen; S., Merlet; T., Farah; C., Guerlin; F., Pereira Dos Santos; N., Le Moigne11; C., Champollion; S., Deville; L., Timmen; R., Falk; H., Wilmes; D., Iacovone; F., Baccaro; Germak, ALESSANDRO FRANCO LIDIA; E., Biolcati; J., Krynski; M., Sekowski; T., Olszak; A., Pachuta; J., Agren; A., Engfeldt; R., Reudink; P., Inacio; D., Mclaughlin; G., Shannon; M., Eckl; T., Wilkins; D., van Westrum; R., Billson. - In: METROLOGIA. - ISSN 1681-7575. - 49:Issue 1A (Technical Supplement 2012)(2012). [10.1088/0026-1394/49/1A/07014]

Final report of the regional key comparison EURAMET.M.G-K1: European Comparison of Absolute Gravimeters ECAG-2011

GERMAK, ALESSANDRO FRANCO LIDIA;
2012

Abstract

During November 2011 a EURAMET key comparison of absolute gravimeters was organized in the Underground Laboratory for Geodynamics in Walferdange, Luxemburg. The comparison assembled 22 participants coming from 16 countries and four different continents. The comparison was divided into two parts: a key comparison that included six National Metrology Institutes or Designated Institutes, and a pilot study including all participants. The global result given by the pilot study confirms that all instruments are absolutely coherent with each other. The results obtained in the key comparison confirm a good agreement between the NMI instruments. Finally, a link to ICAG-2009 shows also that the NMI gravimeters are stable in time.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11696/33287
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? 0
social impact