A trilateral comparison of capsule-type standard platinum resistance thermometers is reported that links the NMIJ and INRIM realisations of the International Temperature Scale of 1990 to the results of the Consultative Committee for Thermometry Key Comparison 2 in the temperature range 13.8033 K to 273.16 K.

Final report on CCT-K2.5: NRC/NMIJ/INRIM comparison of capsule-type standard platinum resistance thermometers from 13.8 K to 273.16 K / Hill, K. D.; Nakano, T.; Steur, PETRUS PAULUS MARI. - In: METROLOGIA. - ISSN 0026-1394. - 52:Tech. Suppl. 03003(2015).

Final report on CCT-K2.5: NRC/NMIJ/INRIM comparison of capsule-type standard platinum resistance thermometers from 13.8 K to 273.16 K

STEUR, PETRUS PAULUS MARI
2015

Abstract

A trilateral comparison of capsule-type standard platinum resistance thermometers is reported that links the NMIJ and INRIM realisations of the International Temperature Scale of 1990 to the results of the Consultative Committee for Thermometry Key Comparison 2 in the temperature range 13.8033 K to 273.16 K.