This technical report describes the characterization procedures performed during the commissioning of the Keysight N5225B Vector Network Analyzer (VNA) at the INRiM High-Frequency Metrology Laboratory. The instrument operates over a continuous frequency range from 9 kHz to 50 GHz. Particular attention is devoted to the evaluation of instrument-related contributions to the uncertainty of scattering parameters measurements, namely, noise floor, trace noise, drift, and linearity. Noise floor, trace noise, and drift are evaluated in accordance with EURAMET Calibration Guide No. 12, whereas linearity is assessed using the Uncompressed Receiver Calibration Method (URCM).
In questo rapporto tecnico si descrivono le operazioni di caratterizzazione eseguite nell’ambito della messa in servizio dell’Analizzatore di Reti Vettoriale (VNA) Keysight N5225B presso il laboratorio di metrologia in alta frequenza dell’INRiM. Lo strumento opera con continuità nella banda di frequenza da 9 kHz a 50 GHz. Particolare attenzione è dedicata alla valutazione dei contributi all’incertezza di misura dei parametri di scattering legati alle caratteristiche dello strumento. In particolare, noise floor, trace noise e drift sono valutati in conformità con il documento EURAMET Calibration guide n. 12 mentre la linearità è valutata mediante il metodo chiamato Uncompressed Receiver Calibration Method (URCM).
Commissioning of the Keysight N5225B Vector Network Analyzer for the INRiM RF&MW Metrology Lab / Oberto, L., Terzi, P.. - (2026).
Commissioning of the Keysight N5225B Vector Network Analyzer for the INRiM RF&MW Metrology Lab
L. Oberto;P. Terzi
2026
Abstract
This technical report describes the characterization procedures performed during the commissioning of the Keysight N5225B Vector Network Analyzer (VNA) at the INRiM High-Frequency Metrology Laboratory. The instrument operates over a continuous frequency range from 9 kHz to 50 GHz. Particular attention is devoted to the evaluation of instrument-related contributions to the uncertainty of scattering parameters measurements, namely, noise floor, trace noise, drift, and linearity. Noise floor, trace noise, and drift are evaluated in accordance with EURAMET Calibration Guide No. 12, whereas linearity is assessed using the Uncompressed Receiver Calibration Method (URCM).I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.