Under EURAMET Project 1687, INRiM and LNE conducted a compatibility study in the resistance range from 1 T to 1 P, comparing the Dual Source Bridge method with the charging-capacitor and sub femto ampere transimpedance amplifier methods. To overcome the detector sensitivity limits, an important bottleneck in low voltage measurements, a validated mathematical model is used. The comparison of three independent methods against a recognized mathematical model derived from published metrological capabilities is presented on a 100 TΩ network-type resistor.
Ultra-high electrical resistance calibration at 100 T addressing the open problems identified in the low voltage measurements / Mihai, I., Patois, E., Capra, P.P., Galliana, F., Piquemal, F., Istrate, D.. - digest of 35th edition of the CPEM, « Conference on Precision Electromagnetic Measurements:(2026). (35th edition of the CPEM, « Conference on Precision Electromagnetic Measurements MADRID 6th to the 11th of September 2026) [10.1109/CPEM68396.2026.11756020].
Ultra-high electrical resistance calibration at 100 T addressing the open problems identified in the low voltage measurements
I. Mihai
Conceptualization
;P. P. CapraMembro del Collaboration Group
;F. GallianaWriting – Original Draft Preparation
;
2026
Abstract
Under EURAMET Project 1687, INRiM and LNE conducted a compatibility study in the resistance range from 1 T to 1 P, comparing the Dual Source Bridge method with the charging-capacitor and sub femto ampere transimpedance amplifier methods. To overcome the detector sensitivity limits, an important bottleneck in low voltage measurements, a validated mathematical model is used. The comparison of three independent methods against a recognized mathematical model derived from published metrological capabilities is presented on a 100 TΩ network-type resistor.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.