During the last WGFF meeting (2012) the question of what kind of facilities can be considered as primary standard for airspeed has been discussed. The final decision about dragging facilities that can be accepted as primary standards at a NMI, from the viewpoint of the CIPM 2009-24 (Traceability in the CIPM MRA) [1], clearly limits the possible use of the INRIM facility for anemometer calibration. LDA measurements of airspeed induced by the arm rotation at the measurement position and a definition of actual capabilities of this facility as primary standard according to the WGFF decision are presented in this paper.
Traceability in anemometry: a rotating arm as airspeed primary standard / Piccato, Aline; Spazzini, PIER GIORGIO; Malvano, R.. - (2013). (Intervento presentato al convegno FLOMEKO 2013 - International Conference on Flow Measurement tenutosi a Paris (F) nel September 24-26, 2013).
Traceability in anemometry: a rotating arm as airspeed primary standard
PICCATO, ALINE;SPAZZINI, PIER GIORGIO;
2013
Abstract
During the last WGFF meeting (2012) the question of what kind of facilities can be considered as primary standard for airspeed has been discussed. The final decision about dragging facilities that can be accepted as primary standards at a NMI, from the viewpoint of the CIPM 2009-24 (Traceability in the CIPM MRA) [1], clearly limits the possible use of the INRIM facility for anemometer calibration. LDA measurements of airspeed induced by the arm rotation at the measurement position and a definition of actual capabilities of this facility as primary standard according to the WGFF decision are presented in this paper.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.