The results of a magnetic and electromagnetic measurement survey are presented, which was carried out around a Magnetic Resonance Imaging device. The investigation is performed with the twofold aim of characterising the field source under the different operating sequences and evaluating the exposure of workers both to the static and time-varying magnetic field. The exposure levels are estimated, as a function of the position of the workers and the field characteristics, and their compliance with the existing and possible new limits is discussed. Critical situations are put in evidence in relation to both the exposure limits and the performance of the available measuring instruments.
Analysis of magnetic and electromagnetic field emissions produced by a MRI device / Borsero, Michele; Crotti, Gabriella; Giordano, Domenico; Zucca, Mauro. - CD volume:(2010), pp. 1-5. (Intervento presentato al convegno 17th Symposium IMEKO TC 4 - Measurement of Electrical Quantities tenutosi a Kosice (Slovakia) nel September 2010).
Analysis of magnetic and electromagnetic field emissions produced by a MRI device
BORSERO, MICHELE;CROTTI, GABRIELLA;GIORDANO, DOMENICO;ZUCCA, MAURO
2010
Abstract
The results of a magnetic and electromagnetic measurement survey are presented, which was carried out around a Magnetic Resonance Imaging device. The investigation is performed with the twofold aim of characterising the field source under the different operating sequences and evaluating the exposure of workers both to the static and time-varying magnetic field. The exposure levels are estimated, as a function of the position of the workers and the field characteristics, and their compliance with the existing and possible new limits is discussed. Critical situations are put in evidence in relation to both the exposure limits and the performance of the available measuring instruments.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.