We present a capacitance build-up method suitable for the determination of the measurement error of a capacitance meter. The method requires only a small number of uncalibrated base capacitors, to be connected in parallel in various combinations, and a single calibrated capacitor, which provides measurement traceability. The outcome of the method is both the determination of the meter error and the calibration of all the base capacitors; it can therefore be also considered a capacitance scaling method. The method's equations, cast in matrix form, express estimates and uncertainties for all the quantities of interest. As an example of application, a commercial LCR meter is calibrated in the ranges of (100-1000) pF and (1-10) nF at 1.6 and 10 kHz, with an accuracy at the level of a few parts in 10(6). The calibration is validated by comparison with an ultrahigh accuracy capacitance bridge.

A Capacitance Build-Up Method to Determine LCR Meter Errors and Capacitance Transfer / Tran, Ntm; D'Elia, V; Callegaro, L; Ortolano, M. - In: IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT. - ISSN 0018-9456. - 69:8(2020), pp. 5727-5735. [10.1109/TIM.2019.2960620]

A Capacitance Build-Up Method to Determine LCR Meter Errors and Capacitance Transfer

Tran, NTM;D'Elia, V;Callegaro, L;
2020

Abstract

We present a capacitance build-up method suitable for the determination of the measurement error of a capacitance meter. The method requires only a small number of uncalibrated base capacitors, to be connected in parallel in various combinations, and a single calibrated capacitor, which provides measurement traceability. The outcome of the method is both the determination of the meter error and the calibration of all the base capacitors; it can therefore be also considered a capacitance scaling method. The method's equations, cast in matrix form, express estimates and uncertainties for all the quantities of interest. As an example of application, a commercial LCR meter is calibrated in the ranges of (100-1000) pF and (1-10) nF at 1.6 and 10 kHz, with an accuracy at the level of a few parts in 10(6). The calibration is validated by comparison with an ultrahigh accuracy capacitance bridge.
File in questo prodotto:
File Dimensione Formato  
08937809.pdf

solo utenti autorizzati

Tipologia: Versione editoriale
Licenza: Non Pubblico - Accesso privato/ristretto
Dimensione 2.46 MB
Formato Adobe PDF
2.46 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Tran2019_TIM.pdf

Open Access dal 22/12/2021

Tipologia: Documento in Post-print
Licenza: Pubblico - Tutti i diritti riservati
Dimensione 1.45 MB
Formato Adobe PDF
1.45 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11696/64794
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 5
  • ???jsp.display-item.citation.isi??? 5
social impact