This paper proposes a method for the measurement of digitizer absolute phase errors, defined as the phase displacement between the digitized output and the input analog waveform. The measurement procedure is based on the use of a phase reference signal (PRS), which is theoretically synchronous with the input analog waveform and that triggers the digitizer sampling clock. From the characterization of the waveform generator phase response and the measurement of the time delay of the digitizer sampling clock with respect to the PRS, an accurate evaluation of the digitizer absolute phase error is obtained. The method has been applied to a high-performance digitizer, measuring the absolute phase errors of two different channels. The expanded uncertainty of the method has been quantified as a few microradians at 50 Hz and 150 μrad at 20 kHz.

Measurement of the Absolute Phase Error of Digitizers / Crotti, Gabriella; Femine, Antonio Delle; Gallo, Daniele; Giordano, Domenico; Landi, Carmine; Luiso, Mario. - In: IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT. - ISSN 0018-9456. - 68:6(2019), pp. 1724-1731. [10.1109/TIM.2018.2888919]

Measurement of the Absolute Phase Error of Digitizers

Crotti, Gabriella;Giordano, Domenico;
2019

Abstract

This paper proposes a method for the measurement of digitizer absolute phase errors, defined as the phase displacement between the digitized output and the input analog waveform. The measurement procedure is based on the use of a phase reference signal (PRS), which is theoretically synchronous with the input analog waveform and that triggers the digitizer sampling clock. From the characterization of the waveform generator phase response and the measurement of the time delay of the digitizer sampling clock with respect to the PRS, an accurate evaluation of the digitizer absolute phase error is obtained. The method has been applied to a high-performance digitizer, measuring the absolute phase errors of two different channels. The expanded uncertainty of the method has been quantified as a few microradians at 50 Hz and 150 μrad at 20 kHz.
File in questo prodotto:
File Dimensione Formato  
Meas_phase_post_print.pdf

accesso aperto

Descrizione: Post Print Version
Tipologia: Documento in Post-print
Licenza: Pubblico - Tutti i diritti riservati
Dimensione 1.41 MB
Formato Adobe PDF
1.41 MB Adobe PDF Visualizza/Apri
Measurement of the Absolute Phase.pdf

non disponibili

Descrizione: Versione Editoriale
Tipologia: Versione editoriale
Licenza: Non Pubblico - Accesso privato/ristretto
Dimensione 2.99 MB
Formato Adobe PDF
2.99 MB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11696/61787
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 26
  • ???jsp.display-item.citation.isi??? 19
social impact