A new reference resistive-capacitive voltage divider is presented in this paper, operating in DC and AC from 10 kHz up to 200 kHz. The ratio error is lower than 0.1% and the phase error is lower than 400 urad (1.375’) at 100 kHz. The divider is stable and repeatable. The scale factor and the phase error repeatability, tested in 24 hours continuous measurements repeated during 15 weeks, show a scale factor variation lower than 80 ppm and a phase error variation lower than 20 urad. This allows one to define a calibration matrix of the device bringing its relative measurement uncertainty lower than 0.5∙10^-3 . The divider calibration has been obtained with a step-up method, which could improve traceability in many laboratories, and is described in the paper. The divider can be embedded in a calibration chain to test the accuracy of voltage transducers in the measurement of the AC voltage ripple, in the presence of a significant DC component, as it occurs in many power electronic systems. The voltage calibration chain is based on the use of the reference divider and a voltage injector, the secondary of which is placed in series with a stable DC source. By means of this calibration system, one of the best commercial voltage transducers on the market has been analyzed. The work also highlights how the proposed voltage calibration chain could be used in perspective in a phantom power system for the development of a DC plus AC power standard.

1 kV Wideband Voltage Transducer, a Novel Method for Calibration and a Voltage Measurement Chain / Zucca, Mauro; Modarres, Mohammad; Pogliano, Umberto; Serazio, Danilo. - In: IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT. - ISSN 0018-9456. - (2020), pp. 1-12. [10.1109/TIM.2019.2912589]

1 kV Wideband Voltage Transducer, a Novel Method for Calibration and a Voltage Measurement Chain

Zucca, Mauro
;
Modarres, Mohammad;Serazio, Danilo
2020

Abstract

A new reference resistive-capacitive voltage divider is presented in this paper, operating in DC and AC from 10 kHz up to 200 kHz. The ratio error is lower than 0.1% and the phase error is lower than 400 urad (1.375’) at 100 kHz. The divider is stable and repeatable. The scale factor and the phase error repeatability, tested in 24 hours continuous measurements repeated during 15 weeks, show a scale factor variation lower than 80 ppm and a phase error variation lower than 20 urad. This allows one to define a calibration matrix of the device bringing its relative measurement uncertainty lower than 0.5∙10^-3 . The divider calibration has been obtained with a step-up method, which could improve traceability in many laboratories, and is described in the paper. The divider can be embedded in a calibration chain to test the accuracy of voltage transducers in the measurement of the AC voltage ripple, in the presence of a significant DC component, as it occurs in many power electronic systems. The voltage calibration chain is based on the use of the reference divider and a voltage injector, the secondary of which is placed in series with a stable DC source. By means of this calibration system, one of the best commercial voltage transducers on the market has been analyzed. The work also highlights how the proposed voltage calibration chain could be used in perspective in a phantom power system for the development of a DC plus AC power standard.
File in questo prodotto:
File Dimensione Formato  
08703141.pdf

accesso aperto

Descrizione: © 20XX IEEE. Personal use of this material is permitted. Permission from IEEE must be obtained for all other uses, in any current or future media, including reprinting/republishing this material for advertising or promotional purposes, creating new collective works, for resale or redistribution to servers or lists, or reuse of any copyrighted component of this work in other works
Tipologia: Versione editoriale
Licenza: Pubblico - Tutti i diritti riservati
Dimensione 5.36 MB
Formato Adobe PDF
5.36 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11696/61231
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 9
  • ???jsp.display-item.citation.isi??? 8
social impact