An interferometric bench with 28 m length has been realized at INRiM (Istituto Nazionale di Ricerca Metrologica). The facility is mainly devoted to calibrating electronic distance-measurement devices and absolute interferometers. Two working operating modes are possible: a fringe-counting mode having an uncertainty of U = root(0.07)(2) +(0.013 .L)(2) mu m, where L is the measured length in meters, and a 'Doppler' mode having an uncertainty of U = root(0.02)(2) +(0.1 .L)(2) mu m, the latter allowing measurements at high carriage speed. These uncertainty budgets do not take into account the effect of the air refractive index, which has been evaluated separately on the basis of the Edlen formula leading to a relative uncertainty contribution equal to 3 x 10(-7). An accurate motor control allows speeds from a few micrometers per second up to 2.7 m s(-1), making the device also suitable for the calibration of speed measurement devices.

Traceable 28 m-long metrological bench for accurate and fast calibration of distance measurement devices / Astrua, Milena; Pisani, Marco; Zucco, Massimo. - In: MEASUREMENT SCIENCE & TECHNOLOGY. - ISSN 0957-0233. - 26:8(2015). [10.1088/0957-0233/26/8/084008]

Traceable 28 m-long metrological bench for accurate and fast calibration of distance measurement devices

ASTRUA, MILENA;PISANI, MARCO;ZUCCO, MASSIMO
2015

Abstract

An interferometric bench with 28 m length has been realized at INRiM (Istituto Nazionale di Ricerca Metrologica). The facility is mainly devoted to calibrating electronic distance-measurement devices and absolute interferometers. Two working operating modes are possible: a fringe-counting mode having an uncertainty of U = root(0.07)(2) +(0.013 .L)(2) mu m, where L is the measured length in meters, and a 'Doppler' mode having an uncertainty of U = root(0.02)(2) +(0.1 .L)(2) mu m, the latter allowing measurements at high carriage speed. These uncertainty budgets do not take into account the effect of the air refractive index, which has been evaluated separately on the basis of the Edlen formula leading to a relative uncertainty contribution equal to 3 x 10(-7). An accurate motor control allows speeds from a few micrometers per second up to 2.7 m s(-1), making the device also suitable for the calibration of speed measurement devices.
File in questo prodotto:
File Dimensione Formato  
banco lungo Macroscale 2015.pdf

non disponibili

Tipologia: Versione editoriale
Licenza: Non Pubblico - Accesso privato/ristretto
Dimensione 1.02 MB
Formato Adobe PDF
1.02 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
LORI final version.pdf

accesso aperto

Tipologia: Documento in Pre-print
Licenza: Pubblico - Tutti i diritti riservati
Dimensione 810.77 kB
Formato Adobe PDF
810.77 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11696/29367
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 7
  • ???jsp.display-item.citation.isi??? 6
social impact